GE Healthcare · MRSigna2224842
The Home Flag on the Longitudinal Drive Belt was not passing through the Home Flag Detect Optical Coupler.
Symptom
Table Needs a Push Moving in and Cradle Unlatches When doing heavy patients the customer had to push the cradle to scan position and in some cases the cradle will come unlatched from the dog house.
Error message
2224842 SRI Hardware Error Unexpected cradle motion. Thu Jun 9 13:59:11 2005 Error: 2224842 Host: MR_PHC Process: NSP File: HPC_PORT_ID:sri Line: 99 SRI Hardware Error: Unexpected cradle motion. This error would occur if the table were moved by hand. Please run SRI diagnostics if that is not the case. Expected Cradle Velocity: -100 mm/sec Measured Cradle Velocity: 189 mm/sec
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR