Pular para o conteúdo principal
GE Healthcare · CT

CT/e, CT Prospeed AI/FI Irix or Linux

Symptom

Scan Aborts with JEDI XG Error. Scout Scan Aborts with JEDI XG Error

Error message

Error: 60-0314H ILR Current time out. The current resonant frequency is lower than expected Error code: 60-0314H ILR Current time out. The current resonant frequency is lower than expected See the long description for the full error code Error code: 60-0314H Message/explanation: ILR Current time out. The current resonant frequency is lower than expected Potential cause: 1. kV control 2. Inverter 3. Current measurement circuit. Recommended action/Troubleshooting guide: Run HV power diagnostics.

Applies to

Product
CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
Component
D3161T / D3162T Venus tube used on CTE (D3161T is NEW non reload version) See Long Description for details X-ray tube D3162T / D3161T , Modified Venus tube Part #: D3162T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MX135 CT TH.1.1 RELOAD FRU: Yes Status: Orderable Comments: FLS - ECONOMY SLIP RING CT TUBE (ALSO CALLED CTE HIGHSPEED) 23 OHM (STATOR RESISTOR VALUE) 11/24/99 AM SOURCE = MEXICO FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA Part #: D3161T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MX135 CT TH.1.1 GREEN FRU: Yes Status: Orderable Comments: This is GREEN tube, order only if New is required , If not please SEE D3162T, ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues