VG getting error 144 Velocity test fail for swivel motion axis

Symptom

VG getting error 144 Velocity test fail for swivel motion axis VG getting error 144 Velocity test fail for swivel motion axis.

Error message

VGH Error #144 (Error 144) Velocity test fail axis = 5,6 Swivel VGH Error #144 (Error 144) Velocity test fail axis = 6 Swivel #2 VGH Error #144 (Error 144) Velocity test fail axis = 5 Swivel #1

Applies to

Product
VGH VG and VG Hawkey and Varicam
Component
2199421 (ASM000309) VG HEAD #2 Swivel Assembly 2199421 (ASM000309) VG HEAD #2 Swivel Assembly See Service Note 43671 for VG Swivel Replacement Procedure Old PN ASM000309 Part #: 2199421 Part Type: GP CONSUMABLES Part Weight: -719992.80 LBR Repairable: No Returnable: No Hazardous Material: N Description: 473-3681-4000,HEAD#2 SWIVEL ASSEM FRU: Yes Status: Orderable *Gantry Gantry Gantry 2199420 (ASM000308) VG HEAD #1 Swivel Assembly 2199420 (ASM000308) VG HEAD #1 Swivel Assembly See Service Note 43671 for VG Swivel Replacement Procedure Old P/N ASM000308 Part #: 2199420 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 15.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 473-3681-3903,HEAD#1 SWIVEL ASSEM FRU: Yes Status: Orderable
Area
Nuclear Medicine