GE Healthcare · MRSigna22488312225754

Scan stops with Analogic RF Amplifier 2 Power Supply Fault error 2248831 due to RF Amplifier.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2248831 Analogic RF Amplifier 2 Power Supply Fault. (same as 2225754 on LX platform) Error: 2248831 Host: BIDMCGEMR2 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2505 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Spin Echo Coil: HD BodyFull Scan: 452/1 Analogic RF Amplifier 2 Power Supply Fault. Analogic Amplifier 2 indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated, Power supply FETS are shorted.->

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
Area
MR

Full cause & step-by-step resolution