GE Healthcare · MRSigna2225472
CWA ID16343. cannot scan - RF amp' not ready
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2225472 The RF amplifier failed to go into the commanded mode. Error: 2225472 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3088 RF Amplifier Error. The RF amplifier failed to go into the commanded mode (head or body), or the RF amplifier changed mode when not commanded to change. Present Mode: 0x60 Requested Mode: 0x40 Please power cycle the RF Subsystem and Reset TPS.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.