Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26010720526010740250567

LightSpeed 5.X (60kW)

Symptom

Gantry does not rotate. Customer noticed that gantry does not rotate.

Error message

Error: 260107205 Axial Motor Control (AMC) logic could NOT clear the FAULT line from the AXIAL Drive See Long Description Tue Jun 6 16:50:06 2006 Error: 260107205 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2605 Tue Jun 6 16:50:07 2006 Host : Table Gantry Processor Ermes # : 260107205 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 2605 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic could NOT clear the FAULT line from the AXIAL Drive. Last Axial Drive Fault will be logged before this message.

Error: 260107402 The interlock between the Chopper Control circuit on the Axial I/F Board and the Drive is open. See Long Description for complete error Wed Mar 26 03:57:26 2008 260107402 2 tgp tgp_subsys AX2fault.c 195 Wed Mar 26 03:57:26 2008 Host : Table Gantry Processor Ermes # : 260107402 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 195 Function : AXIAL CONTROL : Axial Communications Scan Type : Axial Scan: 50567/1/2 %s The Axial AC Drive reported fault code: F:2 Auxiliary Fault The interlock between the Chopper Control circuit on the Axial I/F Board and the Drive is 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P101 - FUSE 20 AMPS 700 VOLTS - Fuse used in Axial Dynamic Brake and HEMRC (HSA - 1.X) and also in HEMRC (2.X-5.X) See LD for some of the locations of this fuse 46-170021P101 - FUSE 20 AMPS 700 VOLTS many pictures below 46-170021P101 - FUSE 20 AMPS 700 VOLTS - Fuse for Axial Dynamic Brake Part #: 46-170021P101 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 20 AMPS 700 VOLTS FRU: Yes Status: Orderable Comments: FLS - FUSE 20 AMPS 700 VOLTS USED ON FAST RHAPSODE GWENN 08992 7/2/97 Fuse for 5.X Axial Dynamic Brake part found in LIGHTSPEED 5.X ILLUSTRATED PARTS LIST DIRECTION 2349187-100, REVISION 14 Chapter 4 Gantry (Pro 16) Section 1 HP100 Vertical Assembly Table 4-17 Fuse for PRE 5.X Axial Dynamic Brake Fuse for HEMRC Rotor controler (HSA-1.X) Fuse for HEMRC Rotor Control (2.x-4.x)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues