Scan Prep Fails with Error Code 2227938 The integer value for rdb_hdr_npasses in the header = 0, and is out of range due…
Scan Prep Fails with Error Code 2227938 The integer value for rdb_hdr_npasses in the header = 0, and is out of range due to a known software problem with 15.0 M4 software.
Symptom
Scan Prep Fails with Error Code 2227938 The integer value for rdb_hdr_npasses in the header = 0, and is out of range. See Long Description Scan Prep Fails with Error Code 2227938 'The integer value for rdb_hdr_npasses in the header = 0, and is out of range." iTrak # 13259480 SPR # MRIhc43266 Scan prep fails with the following errors. But scan prep and scan worked ok after pushing scan button again without any action Tue Feb 17 15:17:39 2009 2227938 4 bay4 NSP MH_MBOX_VRE:ed_rdb_hdr_check 10 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: Gradient Echo Coil: 8CTL12 Scan: 2881 The integer value for rdb_hdr_npasses in the header = 0, and is out of range. Tue Feb 17 15:17:39 2009 2218527 1 bay4 scn MsaScanOpsScrn.m 1037 Scanning has stopped. error please try again
Error message
2218527 Scanning has stopped. Error, please try again 2218527 lx-mr scn MsaScanOpsScrn.m 916 Scanning has stopped. error please try again 2227938 The integer value for rdb_hdr_npasses in the header = 0, and is out of range. See Long Description 2227938 mr-excite NSP MH_MBOX_VRE:ed_rdb_hdr_check 10 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LVshim Coil: BODY Scan: 52973/1 The integer value for rdb_hdr_npasses in the header = 0, and is out of range.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.