GE Healthcare · X-ray R&FDefinium3601277
error code 3601277 ,right cell does not stop.
Symptom
error code 3601277 ,right cell does not stop. error code 3601277 ,right cell does not stop.The left and center cells are ok.Impossible to calibrate rejection of the right cell. The right cell is not working even on the table, or OTS .
Error message
3601277 Recovery Class: ABORT Error Text: Generator error 100-1301;(4): AEC Board com problem
Applies to
- Models
- Definium
- Product
- Definium 8000
- Component
- Cable MIS 11750A ref 2407432 Short-circuit in the cable MIS 11750A ref 2407432
- Area
- X-ray R&F
Full cause & step-by-step resolution