GE Healthcare · MRSigna2223883
VRE Reset Failure Due to ICN.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2223883 VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs See long description Mon Jan 26 08:47:14 2009 Error: 2223883 Host: OVMRI1 Process: NSP File: VREReset.cc Line: 937 Exception Class: Unknown Function: Recon: Data Acquisition VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs.-> Extended Error: The TPS reset failure is caused by the failure to communicate with all configured ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.