LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cradle / IMS on GT1700 / GT2000 would not advance from home and makes grinding noise. See Long Description for more details and sound. Cradle / IMS on GT1700 / GT2000 would not advance from home and makes grinding noise This problem happen after moving a large patient off of the table, or after physically moving the IMS to it’s mechanical out of gantry limit (home). Attachment is the sound the table makes when the cradle move button is pushed. The button was pressed 2 times (2 sounds)
Error message
Error: 260134802 IMS position is out of tolerance. See Long Description error messages Thu Oct 20 08:01:13 2005 Error: 260134802 Host: tgp Process: tgp subsys File: gtcb movechk.cxx Line: 581 Thu Oct 20 08:01:13 2005 Host : Table Sub-System Ermes # : 260134802 Exception Class : Secondary Severity : Pri/Soft File : gtcb movechk.cxx Line# : 581 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 IMS position is out of tolerance. current position(um) : 398728 target position (um) : 0 start position (um) : 398728 velocity (um/s): 100000 time from start (ms) : 300 Error: 260134707 Elevation could not stop with stop command. If table is moving downward, check valve See Long Description for error versions Mon Sep 19 08:02:03 2005 Error: 260134707 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 616 Mon Sep 19 08:02:03 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 616 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 793822 um current posit
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5127285 SHUT OFF VALVE ASSY used on GT1700 and GT2000 table See Long Description 5127285 SHUT OFF VALVE ASSY used on GT1700 and GT2000 table Part #: 5127285 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SHUT OFF VALVE ASSY Positoning GT FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required