VG when completed collimator exchange, gantry GDU display NO collimator, coll.exch no complete, please open the handle t…

VG when completed collimator exchange, gantry GDU display NO collimator, coll.exch no complete, please open the handle to continue

Symptom

VG when completed collimator exchange, gantry GDU display NO collimator, coll.exch no complete, please open the handle to continue

Applies to

Models
Millennium
Product
VGH VG and VG Hawkey and Varicam
Component
5197979 (PCA000492) MCH Board for SPARE PART 5197979 (PCA000492) MCH Board for SPARE PART 5197979 MCH Board is not compatable with VG motion software version 6.0AH Verify VG Gantry Firmware version. VG Gantry Firmware Version must be 6.0P7 . See KOID 153674
Area
Nuclear Medicine

Full cause & step-by-step resolution