GE Healthcare · MRSigna
Options will not load from MOD due to MOD drive.
Symptom
Unable to Load Options from MOD See Long Description Options appear to load from the Guided Install GUI, but they never show installed on the options list. MOD is NOT write protected.
Error message
unix: dksc1d4vol: Device not ready, spinning up See Long Description This is seen in /var/adm/SYSLOG. dksc1d4vol converts to scsi bus 1, device 4, which is the MOD on an Octane system. unix: dksc1d4vol: Device spin up failed, unable to use device -- corrective action necessary See Long Description This is seen in /var/adm/SYSLOG. dksc1d4vol converts to scsi bus 1, device 4, which is the MOD on an Octane system. unix: dksc1d4vol: Device not ready : Not ready to perform command See Long Description This is seen in /var/adm/SYSLOG. dksc1d4vol converts to scsi bus 1, device 4, which is the MOD on an Octane system.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2318598 Sony MOD Drive Model SMO-F551-SD Sony MOD Drive 2318598 (Model #SMO-F551-SD) Sony MOD Drive 2318598 (Model #SMO-F551-SD)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.