GE Healthcare · CTLightSpeed200002352
LightSpeed 4.X - GRE
Symptom
Software shut itself down during Detailed Calibrations after clicking Confirm button. See long description for more information. Software shut itself down during Detailed Calibrations after clicking Confirm button. FE was doing Detailed Calibrations and had accidently let the Start Scan button timeout as system was getting ready to start large phantom cal scans. After the FE clicked Resume and then clicked Confirm, the software shut itself down.
Error message
Error: 200002352 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanmgr see long description for complete error Thu Oct 8 15:25:58 2009 Error: 200002352 Host: vct2 Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanmgr.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.