Symptom

Table will not drive in any direction. When table elevation buttons are pressed the gantry will tilt See Long Description When pressing table down the gantry tilts back. When pressing table up the gantry tilts forward. No errors logged on boot up or during a scan hardware reset. Gantry tilt buttons still work normally.

Error message

Error: 185109 Elevation Interference/Limit switches are INVALID. See Long Description for full error Mon Dec 10 16:56:08 2007 Error: 185109 Host: ETC Process: Control File: etcValidIntfMatrix.c 1.8 7/9/ Line: 96 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 16:56:08:425 Log Series: 45 HW error: Elevation Interference/Limit switches are INVALID. Interference/Limit switches = 0xc1d0. UP LIMIT: 1 DOWN LIMIT: 0 INTF1: 1 INTF2: 0 (0 = NOT active 1 = active)

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-297093P1 AC/DC POWER SWITCH - used on Zeus tables See LD for more info 46-297093P1 HSA-CTI AC/DC POWER SWITCH Part #: 46-297093P1 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: AC/DC POWER SWITCH FRU: Yes Status: Orderable FLS - IBS PART AC/DC TOGGLE SWITCH Comments: FLS - IBS PART AC/DC TOGGLE SWITCH
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.