Pular para o conteúdo principal
GE Healthcare · CTLightSpeed185100

LightSpeed 4.X - GRE

Symptom

Cradle will not move in or out of the Gantry. No c-pulse is being detected from longitudal encoder Cradle will not move in or out of the Gantr . No c-pulse is being detected from longitudal encoder y

Error message

Error: 185100 HW error: A to D conversion error. See long description Thu Apr 29 11:12:20 2010 Error: 185100 Host: ETC Process: Control File: ETCa2d.c 1.5 3/17/92 Line: 160 Function: No System Function Reported Scan: 4837/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 11:12:20:700 Log Series: 32 HW error: A to D conversion error. A to D MUX mask = 0x200 conversion status = 0x8090. Here are the Mux read mask registers values.. /*** MUX Select masks */ #define MUX_NEG_15V 0X0000 #define MUX_CRADLE_POT 0x0100 #define MUX_TILT_POT 0x0200 #define MUX_POS_24V 0x0300 #define MUX_CRADLE_DA_OUTPUT 0x0400 #define MUX_ELVTLT_DA_OUTPUT 0x0500 #define MUX_POS_15V 0x0600 #define MUX_ELVTLT_DA 0x0700

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.