MRIQ Intermittent Noisy and Blurry Images Due to STIF Board.

Symptom

MRIQ Intermittent Noisy and Blurry Images See long description Happens randomly on any coil, any scan technique. See attached for image example. 3 acquisition scan shows 1st acquisition good 2nd a little blurry, 3rd has complete loss of signal.

Error message

2253619 Diagnostic SRF IRF Datapath test detected a checksum error on the sequence bus 2253619 ERROR: EM_DIAG_SRF_IRF_SEQ_CHKSUM_BAD SRF->IRF Datapath test detected a checksum error on the sequence bus. Data on this bus is being corrupted by one of the devices hooked to the bus. Extended Error: During this test, data was sent by the SRF to the 32-bit sequence bus on the MDG backplane, past the STIF board, and into the IRF board. The IRF detected that the data was corrupt. This might be caused by a bad board or a bad backplane. You can isolate the effect of the STIF board by removing that board (power down first) and running the SRF-IRF diag again. The system will not run without the STIF board but the SRF-IRF diag can be run. If the test passes without the STIF the STIF is defective. Otherwise, the problem is in the SRF, IRF, or backplane. Check for bent packplane pins. Re-seating the IRF/SRF/STIF might help. Possible cause(s): SRF Board, IRF Board, STIF Board, MGD Backplane

2253618 SRF->IRF Datapath test detected bad data. Diagnostic: SRF->IRF DP Iteration: 1 of 1 Action: 1 Task ID: 0x7f53f40 Time: 11857 Object: APS Test Task Function: Software/Firmware/Config: N/A SRF->IRF Datap

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required