Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2300008975007423000003036250256000549000

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230000897 Write to scan data disk failed. See LD for more details Tue May 23 07:22:28 2006 Error: 230000897 Host: darc Process: dataacq File: raw_file.cxx Line: 250 Function: Data Acquisition : Scan Data Save Scan Type: Not defined Scan: 50074/1/1 Scan Seq Id: 0 Write to scan data disk failed. Status: -1 Most Recent System Error: OS error : 98 --- Address already in use

Error: 230000030 Unable to write offset view data to the scan file. See LD for complete error Error: 230000030 Unable to write offset view data to the scan file Thu Sep 7 12:25:05 2006 230000030 4 ice dataacq dataacq_save.cxx 669 Function: Data Acquisition : Scan Data Save Scan Type: Axial Scan: 36250/200/2 Scan Seq Id: 256000 Unable to write offset view data to the scan file. Bytes being saved: 200000h Save attempt: 649 Last view buffer? 0 Scan aborted. Wed May 4 16:26:10 2005 Error: 230000030 Host: ice Process: dataacq File: dataacq_save.cxx Line: 720 Function: Data Acquisition : Scan Data Save Scan Type: Helical Scan: 2800/3/1 Scan Seq Id: 549000 Unable to write offset view data to the scan file. OS error : 140 --- Remote I/O error Scan file may be deleted manually.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5114536-11 (replaces 5114536-10) Scan Disk Array, Hard drive with mounting sled, Seagate ST37345LC FW001 See Long Description Part #: 5114536-11 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Scan Disk Array, Hard Drive with mounting sled, Seagate ST373455LC Firmware Revision 001 FRU: Yes Status: Orderable Comments: Per ECR 2049162, 212008898, 10/25/07. This part replaces 5114536-10. See the attachment for further details. Part #: 5114536-10 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Scan Disk Array, Hard drive with mounting sled, Seagate ST373454LC Firmware 002 FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041216 __Set as Critical Part for Product CT_VCT, ms04278, 05Oct14 See this link to service note for details this part being obsolete and changed up to 5114536-11 : http://www.ct.med.ge.com/tech_pubs/2360027/sn/SN07102512a.pdf
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues