Pular para o conteúdo principal
GE Healthcare · MRSigna2258107

Port A on the LPCA has a partial short between pins 2 and 3 of the K column.

Symptom

Any receive only coil connected to Port A immediately generates error 2258107 Coil ID light goes through normal sequence (brief red, then green) then error 2258107 is logged.

Error message

2258107 Host coil validation error Body Transmit Enable mismatch detected Error: 2258107 Host: MR3T Process: coilCK File: coilCK scp.c Line: 1026 Exception Class: Unknown Function: Host: Coil Check Host coil validation error: Body Transmit Enable mismatch detected. Connector: Multicoil A Port Coil: HD Shoulder Array by GE - A/B port

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5111020-50 HDMR2 A plus Bezel and cable harness.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues