Symptom

Customer can not page between scout views when using Show Localizer See Long Description for details Customer can not page between scout views when using Show Localizer with Errors 200501536 , Previous message error was of type UIF State Info Action Request with sequence number 538 and 200105007 Scanner software failed to process message with status 200501520 while Enabling Landmark Wed Apr 5 15:24:48 2006 Error: 200501520 Host: bhicct1 Process: scanRx File: uifIpcMgrScanRx.c Line: 6812 Function: Data Acquisition : OC Processing Received message from service uif_state_info with out of order sequence number, expected 538 received 0 Wed Apr 5 15:24:48 2006 Error: 200501536 Host: bhicct1 Process: scanRx File: uifIpcMgrScanRx.c Line: 6834 Function: Data Acquisition : OC Processing Previous message error was of type UIF State Info Action Request with sequence number 538 Wed Apr 5 15:24:48 2006 Error: 200105007 Host: bhicct1 Process: scanRx File: SubSystemIf.c Line: 650 Function: Data Acquisition : OC Processing Scanner software failed to process message with status 200501520 while Enabling Landmark

Error message

Error: 200501520 Received message from service uif_state_info with out of order sequence number, expected 538 received 0 See LD for complete error Wed Apr 5 15:24:48 2006 Error: 200501520 Host: bhicct1 Process: scanRx File: uifIpcMgrScanRx.c Line: 6812 Function: Data Acquisition : OC Processing Received message from service uif_state_info with out of order sequence number, expected 538 received 0

Error: 200501544 Failure had ocurred while performing rxif_setPatientEntry. see long description for full error. Sat Apr 17 15:35:35 2010 200501544 ct99 scanRx ZeusrpGlobalActions.c Function: Data Acquisition : OC Processing scannerPatientOrientation: Failure had ocurred while performing rxif_setPatientEntry.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.