MRSPT SPT crashes with error 2227071 Due to Octane PCI bit 3 card.

Symptom

MRSPT SPT crashes with error 2227071

Error message

2219712 The executable download of psd failed. PSD waveform generation failed. Error: 2219712 Host: hopgemr1a Process: scn File: RxMgrDownload.m Line: 1184 The executable download of psd failed. PSD waveform generation failed.

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

2227071 Pulse Width must be divisible by the hardware's cycle rate times the pulse's resolution Host: t19 Proc: NSP Error: 2227071 File: AGP:createinstr Line: 1 Pulse Width must be divisible by the hardware's cycle rate times the pulse's resolution. pulse-> rf1as value->3200 cycle rate-> 2 resolution-> 0 Host: b5 Proc: NSP Error: 2227071 File: AGP:createinstr Line: 1 Pulse Width must be divisible by the hardware's cycle rate times the pulse's resolution. pulse-> rf1cfl value->3200 cycle rate-> 2 resolution-> 0 Host: t19 Proc: NSP Error: 2227071 File: AGP:createinstr Line: 1 Pulse Width must be divisible by the hardware's cycle rate times the pulse's resolution. pulse-> rf1as value->3200 cycle rate-> 2 resolution-> 0 Host: b5 Proc: NSP Error: 2227071 File: AGP:createi

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2235744-3 PCI ADAPTER CARD for Octane computer
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.