PROTEUS What needs to be done when the KV Control Board in a Proteus JEDI Generator is replaced?

Symptom

PROTEUS What needs to be done when the KV Control Board in a Proteus JEDI Generator is replaced? The replacement board CANNOT be used as received. It comes loaded with V5.47 software for the CT version JEDI Generator. The JEDI will run with this version of software, but strange things will happen, like transposition of values in the kV, mAs, etc. boxes on the Proteus console screen. There is also a floppy disk that comes with the replacement kV Control Board. This floppy disk is NOT to be used. It contains V4.25 software used with the Silhouette VR version JEDI Generator.

Applies to

Product
PROTEUS XRa
Area
X-ray R&F