Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2601001521793926011200926011823654034260107249

LightSpeed 5.X (80/100kW)

Symptom

Scan stopped with unusual clunk noise and most lights on gantry display went out See Long Description During scanning customer heard a "clunk" noise and scanner stopped. All lights on gantry went out except gantry tilt angle. System was rebooted but now scan hardware fails to reset.

Error message

Error: 260100152 Exposure Interlock Opened Please refer to the long description for further details. Fri Feb 8 18:12:19 2008 Error: 260100152 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 386 Fri Feb 8 18:12:19 2008 Host : Table Gantry Processor Ermes # : 260100152 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 386 Function : System Monitoring : Xray Exposure Interlocks Scan Type : Helical Scan: 17939/3/1 Exposure Interlock Opened

Error: 260112009 Emergency Off button has been pressed See LD for full error Fri May 20 15:06:56 2005 Error: 260112009 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 640 Fri May 20 15:06:56 2005 Host : Table Gantry Processor Ermes # : 260112009 Exception Class : Operator Abort Severity : Secondary File : uif_table_tilt_mgr.c Line# : 640 Function : SYSTEM POWER CONTROL : Emergency Stop Scan Type : None/Unknown Scan: 0/0/0 Emergency Off button has been pressed

Error: 260118236 - (70-0501) DC bus out of range Wed Nov 26 12:56:16 2003 Error: 260118236 Host: orp Process: orp_subsys File: gen_exec.c Line: 1923 Wed Nov 26 12:56:14 2003 Host : Jedi Ermes # : 260118236 Exception Class : Abort Severity : Primar

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2351489 NGPDU 50 amp Table Gantry Contactor Ktg See LD for details P/N 2351489 NGPDU 50 amp Table Gantry Contactor Ktg Part #: 2351489 Part Type: GP CONSUMABLES Part Weight: 3.44 LBR Repairable: No Returnable: No Hazardous Material: N Description: 50A CONTACTOR FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required