MRSPT System Locks Up during SPT Scan Due to Defective IPG.

Symptom

MRSPT System Locks Up during SPT Scan See Long description Almost every attempt at SPT during a PM results in failure. SPT clock keeps counting down, but no scans occur. Issue happens at random times during the SPT scans. Usually requires a complete power cycle to recover. System scans normally for customer, no issues at all.

Error message

2223847 IPG boot status packet is not received by NSP. Mon Mar 15 07:26:40 2004 Error: 2223847 Host: MMCLX_2 Process: NSP File: TardisReset.cc Line: 1594 IPG boot status packet is not received by NSP.

2223847 TPS boot status packet is not received by NSP See Long Description Tue Feb 9 11:47:44 1999 Error: 2223847 Host: MRS2 Process: NSP File: TardisReset.m Line: 1177 TPS boot status packet is not received by NSP.

2223831 TPS loader daemon is not up. Mon Mar 15 07:26:38 2004 Error: 2223831 Host: MMCLX_2 Process: NSP File: TardisReset.cc Line: 1287 TPS loader daemon is not up.

2223846 Download status packet from IPG is not received by NSP. Mon Mar 15 07:26:40 2004 Error: 2223846 Host: MMCLX_2 Process: NSP File: TardisReset.cc Line: 1793 Download status packet from IPG is not received by NSP.

Applies to

Models
Signa, Ovation
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion IPGII 2160818-7 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.