GE Healthcare · CTLightSpeed1270483285
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Fastcals aborts and computer locks up
Error message
Error: 4026 An error was found in the GE System Log. The GE System Log file was renamed and then a new log was created. See Long Description for example 4026 : An error was found in the GE System Log. The GE System Log file was renamed and then a new log was created. Mon Apr 5 16:01:25 2010 Error: 4026 Host: nbhct1 Process: geSyslogd File: geSyslogd.c Line: 2923 An error was found in the GE System Log. The GE System Log file was renamed and then a new log was created. Error: 9 near offset: 128 Original filename: /usr/g/service/log/gesys_nbhct1.log Renamed filename: /usr/g/service/log/gesys_nbhct1.log.1270483285.1.clg
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.