GE Healthcare · CTLightSpeed244586245207260001008244584
LightSpeed 4.X - GRE
Symptom
Invalid or mistached DAS type messages appear when tube warm up is attempted. Invalid or mistached DAS type messages appear when tube warm up is attempted
Error message
Error: 244586 the following hardware subsystems reported a failure when initializing: OBC SS (Coll/Das) See the Long Description for the complete error code Error: 244586 Host: qx-ct Process: fwmgr File: HwSetUp.c Line: 2235 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The following hardware subsystems reported a failure when initializing: OBC SS(Coll/Das) Error: 245207 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning. see long description Mon Apr 3 08:29:12 2006 Error: 245207 ls16 scanmgr validate request.cxx 378 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning. Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long Description to see error example Mon Mar 28 03:39:48 2005 Error: 260001008 Host: OBC Process: HOP File: hop_node_init.c Line: 120 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 03:39:46:170 Log Series: 31 The OBC failed to connect with the DCB (could be the CCB) sub-system in th
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2106993-5 FUSE 20 AMPS 500 VOLTS .3 SECONDS See Long Description for details Part #: 2106993-5 Part Type: GP CONSUMABLES Part Weight: 0.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 20 AMPS 500 VOLTS .3 SECONDS FRU: Yes Status: Orderable Comments: FLS - 20AMPS 500V .3SECOND FUSE U/O COST REDUCED GENERATORS SCPU 65/80 MS12746 PER RSPL 5-97 GELS - MAJ LEAD-TIME LE 99/07/26 PA,FUSE 20A 500V .3 SECONDS
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.