Symptom
Cradle will move a few inches and stop. Cradle will move in a few inches and stop
Error message
Error : 260134617 Cradle clutch lines are disconnected. Cannot control cradle clutch See Long Description for complete error. Mon Nov 17 14:04:02 2008 Error: 260134617 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 1040 Mon Nov 17 14:04:02 2008 Host : Table Sub-System Ermes # : 260134617 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 1040 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle clutch lines are disconnected. Cannot control cradle clutch
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5143974 GTCN bd (replaces 5112471) See Long Description Change-Up Data: Part "5112471" changed to "5143974" Part #: 5143974 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCN SET FRU: Not Available Status: Orderable Comments: source pole/planner code revised per to Global Table transfer Part #: 5112471 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCN BOARD ASSY Positioning GT FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.