Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedLightSpeed230023057

BrightSpeed Select or Delight Select

Symptom

Intermittently, when fastcal is run, the System Health Information shown on the Service Desktop will indicate gantry is not balanced. System does not post any error message on screen during fastcal and fastcal runs and completes ok.

Error message

Error: 230023057 Sensor1 (Inner leg) failure. Repair or replace. See long description for full error Mon May 27 20:45:48 2008 Error: 230023057 Host: msec5x1 Process: dailyPrepRx File: DailyPrepUtils.c Line: 4568 Sensor1 (Inner leg) failure. Repair or replace.

Applies to

Models
BrightSpeed, LightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2347474 PIEZO, H-POWER FIELD BALANCE SENSOR See long description for part comments. Part #: 2347474 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PIEZO,H-POWER FIELD BALANCE SENSOR FRU: Yes Status: Orderable Comments: FLS - PIEZO,H-POWER FIELD BALANCE SENSOR. ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19, 03. MS20544 ON 3/3/03.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.