Symptom

Table elevation is very slow going down See long description Actual elevation speeds are as follows: Fast up = 18 sec, slow up = 34 sec, fast down = 38 sec, and slow down = 90+ sec. Spec elevation speeds are as follows: Fast up or down < 30 sec. Full Range slow < 120 sec.

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5245271 GTCB3 (Global Table Control Board 3) This board replaces 5142177-3 GTCB2 and GTCB part # 5112470 thru 5112470-5, and 5142177. See long description for details Part #: 5245271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCB3 BOARD ASSY FRU: Yes Status: Orderable Comments: added per to ECO2080741 by sso#305010201 in May 8, 2009. Part will be available end of Sep, 2009, please order old p/n before new p/n on shelf.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.