Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260134703903212897741751380

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

When the cradle would advance , the vertical number would change to a random number then jump back to another number causing a stop. Once the resume key was pressed cradle would continue to advance.

Error message

Error: 260134703 Elevation moved without command. See LD for more details Wed Jun 14 12:48:34 2006 Error: 260134703 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 530 Wed Jun 14 12:48:34 2006 Host : Table Sub-System Ermes # : 260134703 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 530 Function : No System Function Reported Scan Type : None/Unknown Scan: 412 Elevation moved without command. If characterize completed appears above, this isnt error. If this appears many times, check potentiometer. Current position = 903212 um Last stop position = 897741 um Tolerance = 5000 um Time since last stop = 751380 msec

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5127774 ELEVATION POTENTIOMETER ASSY-VCT See Long Description for details Part #: 5127774 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION POTENTIOMETER ASSY FRU: Yes Status: Orderable Comments: source pole/planner code revised per to Global Table transfer
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.