Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300008975007423000003036250256000549000

LightSpeed 5.X (60kW)

Symptom

Scanner allows tube warm-up but not Fastcals. Scanner allows tube warm-up but not Fastcals

Error message

kernel: XFS: device md0 - XFS write error in file system meta-data block 0x445c998 in md0w kernel: XFS: device md0 - XFS write error in file system meta-data block 0x445c998 in md0

Error: 230000897 Write to scan data disk failed. See LD for more details Tue May 23 07:22:28 2006 Error: 230000897 Host: darc Process: dataacq File: raw_file.cxx Line: 250 Function: Data Acquisition : Scan Data Save Scan Type: Not defined Scan: 50074/1/1 Scan Seq Id: 0 Write to scan data disk failed. Status: -1 Most Recent System Error: OS error : 98 --- Address already in use

Error: 230000030 Unable to write offset view data to the scan file. See LD for complete error Error: 230000030 Unable to write offset view data to the scan file Thu Sep 7 12:25:05 2006 230000030 4 ice dataacq dataacq_save.cxx 669 Function: Data Acquisition : Scan Data Save Scan Type: Axial Scan: 36250/200/2 Scan Seq Id: 256000 Unable to write offset view data to the scan file. Bytes being saved: 200000h Save attempt: 649 Last view buffer? 0 Scan aborted. Wed May 4 16:26:10 2005 Error: 230000030 Host: ice Process: dataacq File: dataacq_save.cxx Line: 720 Function: Data Acquisition : Scan Data Save Scan Type: Helical Scan: 2800/3/1

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2362873 SDDA (Scan data disk array) used on GOC3. See Long Description for details. The GOC3 to GOC4 upgrade kits are now under 1 part number, 5144058-500 plus the FE will have to also order the DARC2 that is needed by the DAS as a separate order. Part #: 2362873 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SCAN DATA DISK ARRAY - GRE FRU: Yes Status: Orderable Comments: OBSOLETE PART case 8391074 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 05-23-07 ** FE ASSIST ** For CT not PET.. IF no stock of 2362873 SDDA must upgrade to DARC II + ICOM Kit per one of the following part number's * 5144058-500, 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL * 5144058-501, 4,8 MDAS DARC Replacement Kit, JARRELL * 5144058-502, 4 SDAS DARC Replacement Kit, JARRELL Documentaion 5194645-100 Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better (2/19/07 212001261) *** 2ND OPINION NOTES *** Look at the DARC messages log for SCSI disk errors located at /var/log/messages
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.