GE Healthcare · CTLightSpeedVCT260001008260001085
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Customer gets Error message 'DAS gain calibration has not been run since converter boards were changed' while running fast cals. Customer gets Error message 'DAS gain calibration has not been run since convertor boards were changed' while running fast cals.
Error message
Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long Description to see error example Mon Mar 28 03:39:48 2005 Error: 260001008 Host: OBC Process: HOP File: hop_node_init.c Line: 120 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 03:39:46:170 Log Series: 31 The OBC failed to connect with the DCB (could be the CCB) sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS Error: 260001085 Communication error: Message not sent to the DCB. Communications with the DCB are down. CAN message could not be sent. Message code: 11 See Long Description for complete error Sun Sep 5 08:05:00 2004 Error: 260001085 Host: OBC Process: HOP File: hop_config_track.c Line: 164 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:04:45:705 Log Series: 30 Communication error: Message not sent to the DCB. Communications with the DCB are down. CAN message could not be sent. Message code: 11
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2289548 MDAS BACKPLANE/CHASSIS, CTR ASM used on 4/8 slice LS 3.x MDAS See long description for details Part #: 2289548 Part Type: GP CONSUMABLES Part Weight: 7.52 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS BACKPLANE/CHASSIS, CTR ASM FRU: Yes Status: Orderable Comments: FLS - MDAS CHASSIS, CENTER ASM-H3. USED ON WARP3. SUPPLIER IS CT MANF. ADD FOR ADMIN REASONS ONLY. WILL NOT BE SUPPORTED BY CA-GPUNTIL M3 (FW38-40). MSA6724 EJM 05/16/01 FOR CHASSIS REPLACEMENT, ORDER FLEX BOOT KIT PN 2329000THIS KIT CONTAINS 120 FLEX BOOT COVERS. ADDED PER CT SERVICE ENGINEERING (ERWIN SULMA) BY MS20544, CSV ON 1/20/02. SEE FE ASSIST. FLS - FE ASSIST: RETURN TO CT SERVICE ENGINEERING FOR EVALUATION. ATTN: TIM WATKINSGE MEDICAL SYSTEMS, 16800 W. RYERSON RD., NEW BERLIN, WI 53151. PH 262 785-4221.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.