GE Healthcare · CTLightSpeed18241535060
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Table keeps moving down after push buttons are released -document in RFS - critical issue Table keeps moving down after push buttons are released
Error message
Error: 182415 HW error:Elevation drive is NOT stopping within 1500 msec. See long description for full error Wed Sep 10 17:26:25 2008 Error: 182415 Host: ETC Process: Elevation File: emcShutDown.c 1.41 11/13/92 Line: 84 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 35060/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 17:26:26:100 Log Series: 871 HW error:Elevation drive is NOT stopping within 1500 msec. Current Position 4966 (.1 mm) Elev Direction: 1 (0 = Up 1 = Down)
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2100671 CT TABLE GAS SPRING - Used on ALL HSA, CTi, LightSpeed 1.X through 5.X tables Part #: 2100671 Part Type: GP CONSUMABLES Part Weight: 1.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT Table gas spring FRU: Yes Status: Orderable Comments : FLS - REPLACES 46-327064P1 & 46-296891P1 PER PCN 161843 7-20-93 MS4398. NOT ON SSS: SUSPA IS NOT SUPPLIER. TOOL FOR REPLACING GAS SPRING IS 2144261 ************************************************************PART HAS BEEN DE-LISTED AS A HAZARDOUS MATERIAL ITEM. PART IS NOT HAZARDOUS FOR ANY MODE OF TRANSPORTATION. HAZARDOUS PAPERWORK IS NOT NEEDED. 4/99 ************************************************************ ALSO USED ON QXI LIGHTSPEED ALSO USED ON LIGHTSPEED GELS - "NON FLAMMABLE GAS - HAZ IATA DESC. ""ARTICLES, PRESSURIZED, PNEUMATIC "" UN 3164 - CLASS 2.2",GAS SPRING HAZ #8 is the shock
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.