GE Healthcare · MRSigna2248652
E02 error on Airsys Medicool Chiller and Patient Comfort Sensor trip errors.2248652 caused by low coolant level in chill…
E02 error on Airsys Medicool Chiller and Patient Comfort Sensor trip errors.2248652 caused by low coolant level in chiller.Will Not Scan With Patient Comfort Sensor Trip.Add Gradient Coolant.Signa HDx…
Symptom
Will Not Scan With Patient Comfort Sensor Trip. See Long Description Patient comfort sensor trip; cannot scan due to bore sensor indicating temperature is too hot for patient safety. A message "Patient Comfort Sensor Trip" appears in the message box on the left side of the scan page at the start of autoprescan.
Error message
E02 Error on AirSys Chiller Display. The AirSys Medicool chiller control display reads Error E02 as the error code for Low pressure (digital) or high water level. Active if low pressure (suction pressure) is below 0.6Bar or when the water level in the tank is below the high level. (Likely a coolant leak) See Long Description. The AirSys Medicool chiller control display reads Error E02 as the error code for Low pressure (digital) or high water level. Active if low pressure (suction pressure) is below 0.6Bar or when the water level in the tank is below the high level. (Likely a coolant leak) The Air Sys Medicool 10.0 P6 R407C chiller User Manual is here . (see page 35) 2248652 Scan inhibited. Patient comfort sensor trip. See Long Description Mon Apr 5 07:37:02 2004 Error: 2248652 Host: lxmr3 Process: NSP File: SCP:BoreTempSriHWO Line: 999 Scan inhibited. Patient comfort sensor trip. To return to scanning, verify: 1. Patient fan is ON. 2. Scan room temperature is normal. 3. Air flow through the bore is unobstructed. 4. Gradient Cooling fan is normal. 5. BRM heat exchanger is normal. Extended Error: The bore temp that was measured by the SRI has reached the second trip level. Subseque
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2297672 50/50 PROPLYENE GLYCOL MIX GE number 2297672 is premixed with Distilled Water and YELLOW in color. GLYCOL, 1 GALLON/3.8 LITERS. Coolant is the same GE part number in both Ellis and Watts and Medicool chillers and is a mixture of propylene glycol and purified water (50/50 mixture) GE number 2297672 is premixed with Distilled Water and YELLOW in color. GLYCOL, 1 GALLON/3.8 LITERS. Coolant is the same GE part number in both Ellis and Watts and Medicool chillers and is a mixture of propylene glycol and purified water (50/50 mixture)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.