Pular para o conteúdo principal
GE Healthcare · MRSigna20000235612630802331551639

Host-ICN Ethernet Stress Diagnostic failed with error 200002356 due to defective network cable form PC to system cabinet…

Host-ICN Ethernet Stress Diagnostic failed with error 200002356 due to defective network cable form PC to system cabinet network switch.

Symptom

200002356 Host-ICN Ethernet Stress Diagnostic failed with error .

Error message

200002356 Host-ICN Ethernet Stress Diagnostic failed see long Error: 200002356 Host: lx-mr6 Process: hostMonitor File: ssProcStop.c Line: 1147 Exception Class: Unknown Function: Host: System Startup dsserver died unexpectedly, no action being taken. Task ID: 0x0 Time: 1263080233 Object: Host Test Process Function: Software/Firmware/Config: N/A Host-ICN Ethernet Stress Diagnostic failed. [[icn2 MAC=00:30:48:5B:90:E0 SERIAL=1551639]] failed to respond->

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5160775 FRU ICN Ethernet Cable Collector This kit contains: 3 cables between ICN and Ethernet Switch, 1 J63 to Ethernet Switch cable, 1 Ethernet cable 100 ft (30.5m) in length between Host and J63 on System Cabinet, and 1 cable label book. HDx system.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.