Pular para o conteúdo principal
GE Healthcare · CT

ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt)

Symptom

Gantry will not rotate

Error message

ER-102 Gantry Error code 4 see LD ER–102 Gantry Error code 4 0000000X Error Code X = 1: Gantry Request Command Error 2: Gantry Set Timeout in 20 sec 3: Servo Motor Overheat 4: Servo Amp Alarm 5: Gantry is not in System Mode 6: Gantry Rotation Not Response 7: Gantry Rotation Overspeed 8: Gantry Rotation Underspeed 9: Safety Switch Open A: Dynamic Brake ON B: Gantry Initialize Error or Position Error F: GNTINITREQ Timeout For ER-102 code 4 Servo Amplifier showed seven segment LED code 1 for Over current.

Applies to

Product
ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
P9183SB servo amplifier assy - Prospeed see LD Part # P9183SB Part Type: GP CONSUMABLES Part Weight: -1451985.48 KGM Repairable: No Returnable: No Hazardous Material: N Description: SERVO_AMP FRU: Yes Status: Orderable Parts Offered as: Current: P9183SB Harvested: Not Available Exchange: Not Available Comments: ***GPRS DISCOUNTABLE PART 9/08*** 212008582 07-OCT-2008
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues