GE Healthcare · CTLightSpeedVCT200002356200002372200002352
LightSpeed 4.X - GRE
Symptom
During applications start up, "loading data" appears on right monitor and software shuts back down to blue screens See long description All applications processes initialize except the stcIf process pwho >> all processes up except stcif stcIf.init *** NOT RUNNING stcIf Then another pwho shows the scanmgr process down pwho scanmgr.init *** NOT RUNNING scanmgr Until all processes stop running
Error message
Error: 200002356 retroRecon terminated unexpectedly, no action being taken See long description Wed Mar 10 07:29:27 2010 Error: 200002356 Host: redm_ge_ct02 Process: cupMonitor File: ssProcStop.c Line: 979 retroRecon terminated unexpectedly, no action being taken. Error: 200002372 The process retroRecon took too long to start up See long description Wed Mar 10 07:29:23 2010 Error: 200002372 Host: redm_ge_ct02 Process: cupMonitor File: ssProcStart.c Line: 856 The process retroRecon took too long to start up. Error: 200002352 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanRx. See long description for complete error Error: Fri Jan 19 16:02:31 2007 Error: 200002352 Host: act8 Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanRx. Mon Nov 28 10:52:30 2005 Error: 200002352 Host: kmcct2 Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiatied by the death of scanRx.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.