GE Healthcare · CTLightSpeedVCT26010002850002
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanning System Aborted Scan. Scanning System Aborted Scan
Error message
Error: 260100028 Scan Control timed out waiting for the Ready For Views message from the OC. See Long Description for complete error. Thu Apr 13 22:25:49 2006 Error: 260100028 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 1616 Thu Apr 13 22:25:50 2006 Host : Table Gantry Processor Ermes # : 260100028 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 1616 Function : DATA ACQUISITION : Scan Data Acquisition Scan Type : Axial Scan: 50002/1/48 Scan Control timed out waiting for the Ready For Views message from the OC. This message is sent from the OC once offset collection has completed. Possible causes: trigger failure, Data Collection failed to receive data, software error. TMO: 5000 msec .
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible - used On GOC6 Disk Array See Long description: Part #: 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible used on GOC6 Disk array ( 5308759 ). Part #: 5308760 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible with 5308759 FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 7/15/08. ** FE ASSIST ** This is a single Hard Drive used in the High Speed Disk Array on the GOC6 Console CT 750 HD and VCT (212001261 23-Jan-2009) ST32050310NS = 250 GB SATA II Hard Disk Drive
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.