Pular para o conteúdo principal
GE Healthcare · MRSigna2249432

Unable to Boot Applications Software with TPS Download Failure due to SCP2 Board in MGD Chassis.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2249432 A PCI error occurred. The PCI System Error SERR was asserted Jun 17 Tue 11:02:15 2249432 diagCoord excLog.c 1338 Task ID: 0x6eca730 Time: 7430 Object: AGP Exception Class: Unknown Function: Software/Firmware/Config: N/A A PCI error occurred. The PCI System Error SERR was asserted. Harrier Err Enable Reg: 0xFFF0E000; Harrier Err Status Reg: 0x04000000. Harrier Err Address Reg: 0x00000000; Harrier Err Attribute Reg: 0x0000. Additional information in /usr/g/service/log/vxPciLg-1cda and /usr/g/service/log/vxPciSum-1d04

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2294300-12 SCP2 Board See Long Description 2294300-12 SCP2 Board. FEMC Parts Comments states as follows: If no stock then 2294300-19 is a suitable substitute. 212002950 February 8, 2010.
Area
MR