Media archiving problems, MOD’s need to run data recovery consistently
Symptom
Media archiving problems. MOD’s need to run data recovery consistently with intermittent Checksum failure See long description Media archiving problems, MOD’s need to run data recovery consistently with intermittent arc_attachDMOD:Checksum failure Possible loss of patient data due to corrupted images after the exams are restored. see below for full details. Failed status 255 for MID:1762902584::2003:01:09:21:26:19GMT:CST Fri Jan 10 17:02:16 2003 Recovery: Recovery took 38.4743 seconds OCT 15 DICOM issue... MESSAGE from Process 4061>> Tue Oct 15 11:25:37 2002 (ArchiveDevice.c:3303) Attach in progress Tue Oct 15 11:25:40 2002 RetrieveDicomDirRec: grp: 20 elm: 32 DS or AN string is formatted incorrectly Tue Oct 15 11:25:40 2002 arc_attachDMOD:Checksum failure Tue Oct 15 11:
Error message
arc_attachDMOD:Checksum failure in arslog in directory /usr/g/ctuser/logfiles Recovery Failed status 255 for MID:1762902584::2003:01:09:21:26:19GMT:CST Fri Jan 10 17:02:16 2003 Recovery: Recovery took 38.4743 seconds cd /usr/g/ctuser/logfiles arslog arc_attachDMOD:Checksum Sometimes the recovery fails after 40 seconds but upon reattaching the MOD the no corrupt media is detected. "Recovery Failed status 255 for MID:1762902584::2003:01:09:21:26:19GMT:CST Fri Jan 10 17:02:16 2003 Recovery: Recovery took 38.4743 seconds OCT 15 DICOM issue... MESSAGE from Process 4061>> Tue Oct 15 11:25:37 2002 (ArchiveDevice.c:3303) Attach in progress Tue Oct 15 11:25:40 2002 RetrieveDicomDirRec: grp: 20 elm: 32 DS or AN string is formatted incorrectly Tue Oct 15 11:25:40 2002 arc_attachDMOD:Checksum failure Tue Oct 15 11: arslog: END MESSAGE from Process 5997>> Fri Jan 10 16:06:31 2003 (arsjob.c:341) [Arserver]> init of arsjob connected to NETserver MESSAGE from Process 5997>> Fri Jan 10 16:06:31 2003 (ArchiveDevice.c:4615) [Arserver]> Resetting ARS SONYSMO-F551 - $Id: @(#)ArchiveDevice.oN32 vobadm Thu Jun 13 03:16:41 CDT 2002 /view/IRIX646.5__build__R2.9j $ MESSAGE from Process 5997>> Fri Jan 10 17:
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2300169-2 Octane II Dual CPU See long description for details .Part #: 2300169-2 Part Type: GP REPAIRABLES Part Weight: 27.28 LBR Repairable: Yes Returnable: Yes Hazardous Material: Y Description: OCTANE II FRU, CT DUAL PROC UNIT FRU: Yes Status: Orderable Comments: FLS - OCTANE 2, DUAL PROCESSOR REPAIRABLE FRU. USED ON CT'S HELIOS 3. MS26561 DSM 7/9/01 SUPPLIER IS SGI. ALL FAILED UNITS MUST BE RETURNED TO RDC INMILWAUKEE. MS20544, CSV ON 1/21/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.