GE Healthcare · CTLightSpeedVCT290014004
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Images Recon is very slow, some images taking 20 minutes to complete
Error message
Do Not Use - use one of the other Error: 290014004 Unable to verify DARC system startup See Long description for complete error NOTE: This is error does not mean you need to replace the darc. Troubleshooting is necessary This error comes in different contexts per the below Mon Dec 12 08:00:11 2005 Error: 290014004 Host: ct99 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 172 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting for ping darc response (exit status: 8) Fri Dec 8 14:42:12 2006 Error: 290014004 Host: kghcateqxi Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - scan disk array is non-recoverable (exit status: 22) Fri Mar 16 09:13:03 2007 Error: 290014004 Host: ctflsh Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting for rsh darc response (exit status: 9) Fri Mar 14 05:49:37 2008 Error: 290014004 Host: nvgect01 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 180 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5183547-38 HP8400 with 4GB Memory See Long Discription for Details Part #: 5183547-38 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: GEHC MI and CT Integrated Buy Level PC Workstation. HP8400 with 4GB Memory FRU: Yes Status: Orderable Comments: Per ECR 2050682 & GOC6 RSPL, 212008898, 7/14/08. ** FE ASSIST ** This part is not qualified for PET systems, CT only This part requires a Software Load from Cold LFC during replacement. For all CT and RT Products, verify you have the latest software and Service Pack for your system before trying to install this part. The software versions listed below are a reference, refer to the latest revision of the software Load Cold Document for your system for more information. For 4,8,16 slice CT HPower and RT4 slice systems, order 07MW11.10, For 4,8,16 slice CT Brightspeed systems, order 07BW08.4 For 16 slice RT systems, 08BW20.4 For 32, 64 slice GOC5 based systems, order 07MW18.4, For 32, 64 slice GOC6 based systems, order 08MW33.2 Future versions of the software releases noted above will are also compatible with this part. (212001261 10/21/08) 5255299-20 VDIG VCT Data Aquisition and Image Generator Computer used on GOC6 5255299-20 VDIG VCT Data Aquisition and Image Generator Computer used on GOC6 Part #: 5255299-20 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDIG - VCT Data Aquisition and Image Generator Computer Assembly FRU: Yes Status: Orderable Part #: 5255299-20 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDIG - VCT Data Aquisition and Image Generator Computer Assembly FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.