Symptom

Fastcal is failing and partial ring artifact seen on images. See Long Description for a picture of the artifact Fastcal is failing and partial ring artifact seen on images.

Error message

Error: 260007937 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Long Description for full error. Tue Dec 28 09:41:32 2004 Error: 260007937 Host: orp Process: orp_subsys File: brdflt.c Line: 82 Tue Dec 28 09:41:31 2004 Host : Das Ermes # : 260007937 Exception Class : Soft Severity : Sec/Soft File : brdflt.c Line# : 82 Function : No System Function Reported Scan Type : None/Unknown Scan: 50444/1/0 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Following Error Message for Converter Card List Board : 87 ASIC/Channel : 0x012c Stage : 1 Board Channel : 41

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2254888-2 TWIN 8 slice Elastomers BOX OF 25 ( DUROMETER ) 2254888-2 TWIN 8 slice Elastomers BOX OF 25 ( DUROMETER ) The Gdas uses the twin elastomer for both the 8 slice and 4 slice systems. only 16 slice systems use interposers 2254888-2 TWIN 8 slice Elastomers BOX OF 25 ( DUROMETER ) The Gdas uses the twin elastomer for both the 8 slice and 4 slice systems. only 16 slice systems use interposers Part #: 2254888-2 Part Type: GP CONSUMABLES Part Weight: 0.24 LBR Repairable: No Returnable: No Hazardous Material: N Description: 8 SLICE 70 DUROMETER FRU: Yes Status: Orderable Comments: FLS - B TWIN ELAST. (25PCS) IN CONTAINER IF YOU ORDER QTY ONE OF THIS PART YOU WILL GET BOX OF 25 MS13452 PER AVNET: THEY NO LONGER SUPPLY THIS ITEM. MS27911 DDI, CURRENT SUPPLIER FOR CTM MS13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues