GE Healthcare · MRSigna222466622522932252305
Unable to scan body with TR Driver faults due to Feed Through Connector at J4 on the Pen Panel - RF Transmit Head Heliax…
Unable to scan body with TR Driver faults due to Feed Through Connector at J4 on the Pen Panel - RF Transmit Head Heliax Connection.
Symptom
Unable to Scan Body with TR Driver Faults Unable to scan body with TR Driver faults
Error message
2224666 Auto Prescan failed, Driver Module Error Occurred. Wed Apr 21 20:00:24 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1334 Auto Prescan failed, Driver Module error occurred. 2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.-> 2252305 TR driver fault detected - open circuit Source of fault is Head T/R on Connector A Value 370 mA, limit 500 mA
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 46-221865P1 CONNECTOR (N BULKHEAD, 50-OHM, JACK-JACK BULKHEAD ADAPTOR) See Long Description This is the RF head transmit bulkhead feed-through "N" connector used on the Penetration Panel to connect to the head heliax RF cables on both sides of the Pen Panel. The FEMC description is simply "CONNECTOR", but in the engineering database (ePDM) the part description is "UG-30, N BULKHEAD, 50-OHM, JACK-JACK PRESSURIZED BULKHEAD ADAPTOR, .546~ X .641~ PANEL HOLE. TROMPETER & BEND". An illustration of this part is shown below.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.