Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT245051173001178517244555178575244518

LightSpeed 2.X (Plus, QX/i) - GRE

Symptom

Customer gets Error message 'DAS gain calibration has not been run since converter boards were changed' while running fast cals. Customer gets Error message 'DAS gain calibration has not been run since convertor boards were changed' while running fast cals.

Error message

Error: 245051 Error verifying that Firmware Manager is ready for scanning. See LD for details Wed Oct 4 15:30:39 2006 Error: 245051 Suite: SKCT Host: SBC0 Process: scanmgr File: start_scanning.m 1.34 Method: start_scanning() Line: 306 Function: Data Acquisition : SBC Processing Scan: 6019/1/1 Type: Scout Scan Seq: 173001 Error verifying that Firmware Manager is ready for scanning

Error: 178517 Failure writing to socket connected to controller OBCR. No route to host See LD for complete error Fri Oct 8 07:49:13 2004 178517 4 ct03 fwmgr NBSSocket.c 690 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failure writing to socket connected to controller OBCR. OS error : 113 --- No route to host May also see the following OS error: OS error : 101 --- Network is unreachable

Error: 244555 The OBC controller has failed to reset See Long Description for the full error Error: 244555 Host: ct01_sbc Process: fwmgr File: HwSetUp.c Line: 724 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The OBC controller has failed to reset.

Error: 178575 Failed to R-HARD the following controllers: STC OBC ETC. See long descripti

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues