GE Healthcare · MRSigna22238472223625108193444522238172223621
TPS Boot Status Packet is Not Received by NSP Errors Due to Medcam board.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2223847 TPS boot status packet is not received by NSP See Long Description Tue Feb 9 11:47:44 1999 Error: 2223847 Host: MRS2 Process: NSP File: TardisReset.m Line: 1177 TPS boot status packet is not received by NSP. 2223625 Send packet opcode:814 seqnum:9 rev:2 type:1 length:328 req:psc_major resp:ps_exec status:0 failed. During TPS Reset: 1081934445 1 1 Wed Apr 14 09:20:45 2004 2223625 4 CARD psc mrmailmgr.c 791 Send packet opcode:814 seqnum:9 rev:2 type:1 length:328 req:psc_major resp:ps_exec status:0 failed. 2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 2223621 Internal Software Error. Can''t callback on undefined NMRID:ps_exec. During TPS Reset: 1081934445 0 1 Wed Apr 14 09:20:45 2004 2223621 4 CARD psc mrmail_connect_mgr.c 238 Internal Software Error. Can''t callback on undefined NMRID:ps_exec. 2223621 Internal Software Error. Can't callback on undefined NMRID RtpServer 2223621 OMLXHI2 RtpHost mrmail connect mgr.c 232 Internal Software Error. Can't callback on undefined NMRID:RtpServer.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion 40MHZ MEDCAM W/AMI ASIC, 2292662-11 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.