Pular para o conteúdo principal
GE Healthcare · MRSigna22238472223625108193444522238172223621

TPS Boot Status Packet is Not Received by NSP Errors Due to Medcam board.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2223847 TPS boot status packet is not received by NSP See Long Description Tue Feb 9 11:47:44 1999 Error: 2223847 Host: MRS2 Process: NSP File: TardisReset.m Line: 1177 TPS boot status packet is not received by NSP.

2223625 Send packet opcode:814 seqnum:9 rev:2 type:1 length:328 req:psc_major resp:ps_exec status:0 failed. During TPS Reset: 1081934445 1 1 Wed Apr 14 09:20:45 2004 2223625 4 CARD psc mrmailmgr.c 791 Send packet opcode:814 seqnum:9 rev:2 type:1 length:328 req:psc_major resp:ps_exec status:0 failed.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2223621 Internal Software Error. Can''t callback on undefined NMRID:ps_exec. During TPS Reset: 1081934445 0 1 Wed Apr 14 09:20:45 2004 2223621 4 CARD psc mrmail_connect_mgr.c 238 Internal Software Error. Can''t callback on undefined NMRID:ps_exec.

2223621 Internal Software Error. Can't callback on undefined NMRID RtpServer 2223621 OMLXHI2 RtpHost mrmail connect mgr.c 232 Internal Software Error. Can't callback on undefined NMRID:RtpServer.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion 40MHZ MEDCAM W/AMI ASIC, 2292662-11 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues