Pular para o conteúdo principal
GE Healthcare · CTDiscoveryVCT260134805

Table will not move down; tape switch was either contacted or not present

Symptom

Table will not move down; with tape or touch switch error See Long Description for switch locations. Table will not move down; tape switch was either contacted or not present The VCT table has a large number of touch sensors that can produce a variety of error codes.

Error message

Error: 260134805 IMS tape switch IN (rear) isn't present See Long Description for more Details Thu Nov 10 08:55:28 2005 Error: 260134805 Host: tgp Process: tgp_subsys File: gtcb_imc_isrfunc.cxx Line: 163 Thu Nov 10 08:55:28 2005 Host : Table Sub-System Ermes # : 260134805 Exception Class : Secondary Severity : Pri/Soft File : gtcb_imc_isrfunc.cxx Line# : 163 Function : No System Function Reported Scan Type : None/Unknown Scan: 405/2/0 IMS tape switch IN (rear) isn't present. Position is 84 um.

Applies to

Models
Discovery, VCT
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5164938 Tape Switch (208mm) used as IMS Tape Switch IN (rear) replaces 5127462-2 - used on GT2000/GT1700. See long description for details Part #: 5164938 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TAPE SWITH 4 ASSY Positioning GT FRU: Yes Status: Orderable Comments: added per to GT SPCN. source pole/planner code revised per to Global Table transfer Note: the little plastic screw for this part is U0087AA The Latest info is in this table. Switch Type Location Part Number (Latest) Switch Description in Block Diagram Tape Switch Part Number Description Switch Description in gesyslog Purpose Tape Switch TOP COVER FRONT ASSY 5127465-3 T-Sensor_Elev1 T-Sensor_Elev2 T-Sensor_Elev3 T-Sensor_Elev4 5133731 TAPE SWITCH 2 (185mm) Elevation touch switch1 (tape switches on underside of top front cover) Detection of clog in elevation down 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch1 (tape switches on underside of top front cover) Detection of clog in elevation down IMS LEFT SIDE COVER 5127463-2 T-Sensor_IMS3 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) Detection of pinching between IMS and column in elevation down T-Sensor_Elev10 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover) Detection of clog in elevation down IMS REAR UPPER COVER 5127462-2 T-Sensor_IMS1 T-Sensor_IMS2 5164938 TAPE SWITCH 4 (208mm) IMS Tape Switch IN (rear) Detection of pinching between IMS and column in elevation up IMS RIGHT SIDE COVER 5127464-2 T-Sensor_IMS4 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) Detection of pinching between IMS and column in elevation down T-Sensor_Elev11 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover) Detection of clog in elevation down Limit Switch GT REAR TOUCH SWITCH 5125112 T-Sensor_Elev7 T-Sensor_Elev8 T-Sensor_Elev9 N/A N/A Elevation touch switch3 (switch on underside of top rear cover) Detection of clog in elevation down TOUCH SENSOR R ASSY 5127469 T-Sensor_Elev5 N/A N/A Elevation touch switch2 (switches on underside of top side cover) Detection of clog in elevation down TOUCH SENSOR L ASSY 5127470 T-Sensor_Elev6 N/A N/A Elevation touch switch2 (switches on underside of top side cover) Detection of clog in elevation down U0087AA GT200/1700 plastic screw See long description for details Part #: U0087AA Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCREW FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues