GE Healthcare · CTLightSpeedVCTDiscovery290009011
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Customer complains that image recon time is slower than normal.
Error message
Error: 290009011 Timeout waiting for abort response from the IG. Will reset the IG (1,2,3) See Long Description for more details Error: 290009011 Timeout waiting for abort response from the IG. Will reset the IG. (1,2,3) Tue Aug 15 13:00:05 2006 Error: 290009011 Host: darc Process: recon_control File: recon_control_cb.cxx Line: 1066 Timeout waiting for abort response from the IG. Will reset the IG. This could be for IG 1,2,3 or VIG 1,2,3
Applies to
- Models
- LightSpeed, VCT, Discovery
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5255299-30 - VRB - pronounced VeRB - Volume Reconstruction Box - USED on GOC6 Console - CT750 HD See Long Description for Details. 5255299-30 - VRB - pronounced VeRB - Volume Reconstruction Box - USED on GOC6 Console - CT750 HD Part #: 5255299-30 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VRB - pronounced VeRB - Volume Reconstruction Box FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required