LightSpeed 4.X - GRE
Symptom
Less image space available than normal. See long description for examples. Less image space available than normal, even after removing all exams in Image Works. Example: H3 or GOC1 console should have 20,000 (512 images) available on a clean system with an Octane cpu with two 9 GB disk drives. But with the replacement of the 9GB drives with 18 and 73GB drives, image space may increase up to 130,000 (512 images) depending on software. Example: GOC2 console should have 100,000 (512 images) available on a clean system with an HP-8000 cpu with one 18GB disk and one 73GB disk. Example: GOC3 or 4 console should have 200,000 (512 images) available on a clean system with an HP-8200 cpu with one 36GB disk and two 73GB disks. Data above is approximate, but can be used to determine if there is a problem with disk space.
Error message
Buffer I/O error on device md0, logical block 78013824 seen on HP or DARC in /var/log/messages See LD for complete error Buffer I/O error on device md0, logical block 78013824 cd /var/log tail –1000 messages | more Look for any disk drive errors like these below in the HP or the DARC messages log: 16:13:48 ctcwct01 kernel: Buffer I/O error on device md0, logical block 26881685 Aug 2 12:11:33 darc kernel: Buffer I/O error on device md0, logical block 78013824 IN /var/log/messages: kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error See long description for details IN /var/log/messages: uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Open unix shell and login as root cd /var/log cat messages | grep Error Look for any errors similar to these disk drive errors shown below: Mar 21 15:49:38 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:43 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:48 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:53 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:59 uact03 kernel: Info
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.