GE Healthcare · MRSigna2225636
Burn Smell coming from RF Cabinet and 2225636 ERBTEC RF Amplifier fault 50 due to Erbtec Amp.
Symptom
Burn Smell coming from RF Cabinet
Error message
2225636 ERBTEC RF Amplifier fault 50. 100V power supply was less than 85V one second after 100V start relay closed 2225636 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3060 ERBTEC RF Amplifier fault 50.100V power supply was less than 85V one second after 100V start relay closed. Possible defective Power Control Board, AC Switching Module, or excessive loading on power supply bus.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion Erbtec 20KW RF Amplifier 46-271962P1 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution