GE Healthcare · MRSigna2249118

The SRI2 was found to have the wrong DIP-switch setting.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2249118 SRI 2 Hardware Configuration Conflict. The SRI 2 System Configuration Dip Switch is set to 0x1, but should be set to 3 (0011 binary) for a Cylindrical System

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR