GE Healthcare · MRSigna2252310
Driver Module not calibrated.
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252310 DD Driver Fault Detected - Open Circuit See Long Description for extended error message. Fri Oct 22 19:03:04 2004 Error: 2252310 Host: PRD_GE_MR03 Process: NSP File: SCP:DCB Line: 1429 DD Driver Fault Detected - Open Circuit. Source of Fault is BD2 (Dynamic Disable Boards at Front of Body Coil). Value = 3644 mV Limit: 2177 mV-> Extended Error: This usually indicates that a cable somewhere in the circuit path is not connected or one of the load diodes has failed open. Inspect the failing path. Use TR DD MC System Path diagnostic to troubleshoot. In HD/HDx Systems, BD1 is the Dynamic Disable Board at Front of Body Coil BD2 is the Dynamic Disable Board at Front of Body Coil BD3 is the Dynamic Disable Board at Back of Body Coil BD4 is the Dynamic Disable Board at Back of Body Coil In 20.0 and later systems, try running the TR DD System Path diagnostic to troubleshoot.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.